Fuji gas analyzer

Audience: operators recording gas concentrations from a Fuji ZP-series NDIR analyzer, and whoever calibrates it. Scope: capa's fujilib adapter — [devices.params] fields, wide-row emission, the validity state on every reading, the settings and calibration command surface, and what the readings can and cannot be used for.


At a glance

Adapter id capa.devices.fuji
Sibling library fujilib
Real adapter capa.devices.fuji
Sim adapter capa.devices.sim.fuji_sim
Resource scheme serial:<port> — one analyzer per port
Channel binding fuji_channel
Emission shape wide_row — one row per poll, every channel in it
Default poll rate 1 Hz
Declarable settings Per gas: response time, range method, range; output hold and hold mode — see Device settings

Supported hardware

Fuji Electric ZP-series NDIR gas analyzers over Modbus RTU (RS-485, fixed 38400 8-N-1). The ZPA is the model capa has been run against; it is also sold as the CAI (ENVEA) ZPA. Which models share the register map, and what has been checked on hardware, is owned by fujilib.

Two limits belong on this page rather than in a footnote:

Not validated for oxygen-consumption calorimetry

The O2 value over Modbus is the analyzer's display value. It arrives in steps of 0.01 vol% and has passed the analyzer's response-time filter. It has not been compared against the analyzer's analog output, and it is not validated for heat-release-rate calculation. The cone profile's oxygen group still takes an analog input only; a fuji_channel O2 channel is recorded beside it, not in place of it.

The analyzer does not say when it is warming up

After power-on the readings are far off for about a minute — CO2 has been seen at more than twice its range — and no status flag says so. Follow the warm-up time in the analyzer's manual before trusting a run. capa marks readings settling only where it can know something happened: see Warm-up and settling.

Configuration

configs/hardware/fuji_real.toml (excerpt)
[[devices]]
name = "analyzer"
adapter = "capa.devices.fuji"

[devices.params]
port = "COM8"
address = 1
rate_hz = 1.0
auto_reconnect = true
snapshot_period_s = 30.0

[devices.params.channel_map]
CH1 = "co2"
CH2 = "co"
CH3 = "o2"

FujiAdapterParams is extra="forbid".

Key Default Notes
port required COM8, /dev/ttyUSB0. The link is fixed at 38400 8-N-1.
address 1 Station number, 1–31, as set on the analyzer's front panel.
channel_map required The gas on each analyzer channel, CH1–CH12. Asserted by you: see The channel map. Only channels named here can be bound.
rate_hz 1.0 Up to 5. One poll is two Modbus transactions and takes about 0.12 s; the analyzer's own filter makes more than a few polls a second pointless.
timeout_s 0.5 Per-reply timeout.
snapshot_period_s 30.0 Health-ping cadence. The snapshot carries the analyzer's settings, the range method of each measured channel among them (ch3_range_method).
auto_reconnect true A connection failure does not end the stream: every tick of the outage is an error row and the port is reopened on a back-off. Toggles SUPPORTS_AUTO_RECONNECT.
overflow "block" "block" or "drop_newest".
options [] Analyzer options you assert are fitted, e.g. ["auto_calibration", "auto_zero"] for a unit whose calibration gases are plumbed through its valve contacts. Without them the automatic-calibration commands are refused.

In the Setup tab's device form channel_map is a JSON line, {"CH1": "co2", "CH2": "co", "CH3": "o2"}. Discovery pre-fills it with what the analyzer's type code suggests.

Channels and emission shape

Every poll reads all of the analyzer's channels and its status, and becomes one fujilib Sample. The native row is preserved into device_records/fuji.parquet via fujilib.sample_to_row: per channel the value, unit, gas, validity state and hold / calibration / error flags; then the analyzer-level errors and alarms; then error_type and error_message.

The only supported binding source is fuji_channel:

[[channels]]
name = "gas.o2"
kind = "gas_concentration"
unit = "percent"
derived_unit = "percent"
plot_group = "gases"

[channels.source]
source = "fuji_channel"
device = "analyzer"
channel = "CH3"

[channels.calibration]
kind = "identity"
input_unit = "percent"
output_unit = "percent"

field picks what the channel carries:

field Value
"value" (default) The concentration, in the analyzer's unit for the channel's current range.
"valid" 1.0 while the reading is live, 0.0 while it is not. No sample while the analyzer does not report validity.

The validity state

Each reading has one state, and it travels on the ChannelSample as status. Only ok is a live measurement:

status Meaning
ok A live reading.
hold The outputs are held; the value is frozen.
calibrating A manual zero or span is under way on the channel.
auto_calibration The analyzer's automatic calibration is running.
analyzer_error The analyzer reports an instrument error (errors 1, 2, 3, 10).
channel_error The channel reports an error (errors 4–9).
settling Read within 90 s of a reconnect. See Warm-up and settling.
source_invalid A derived channel whose source channel or O2 channel is not valid.
unknown The status was not read, or the value did not decode.

When several apply, the first in this order wins: analyzer error, channel error, calibrating, auto calibration, hold, source invalid, settling.

The raw value is kept in every state: a held or calibrating reading is recorded as read, with its state beside it. capa stores status in scalars.parquet, but plots, alarms and procedures do not read it. To act on validity, bind a second channel with field = "valid" and watch that. plot = false keeps it off the Run tab's plots, where a flag that sits at 1 would only draw a flat line; it is still recorded and shown in the Numerics dock. Add one where an alarm or a procedure needs it. The shipped fuji_real.toml declares none.

[[channels]]
name = "gas.o2_valid"
kind = "gas_concentration"
unit = "dimensionless"
derived_unit = "dimensionless"
plot = false

[channels.source]
source = "fuji_channel"
device = "analyzer"
channel = "CH3"
field = "valid"

[channels.calibration]
kind = "identity"
input_unit = "dimensionless"
output_unit = "dimensionless"

Failed polls

A poll that fails still produces a row: the same columns, the readings empty, error_type and error_message filled. No ChannelSample is derived from it, so a gap in scalars.parquet is a gap in the data, and device_records/fuji.parquet says why. With auto_reconnect = false any failed poll ends the stream.

The unit check

A range change at the analyzer can change a channel's unit (vol% on one range, ppm on the other). The adapter compares every reading's unit with the unit its capa channel declares. On a mismatch it quarantines the channel for the rest of the run — no further samples — and emits one unit_mismatch event, rather than record values wrong by a factor of 10,000. Fix the channel's unit or the analyzer's range; the next run starts clean.

See channel bindings for the selector contract.

Events

Each change is one DeviceEvent in the run's events.sqlite:

kind When
comm_lost / comm_restored The first failed poll of an outage, and the first good one after it, with the outage's length and when settling ends.
instrument_error The analyzer's instrument errors change; the codes are in the event.
hold Output hold comes on or goes off.
calibration A zero or span ends — made at the front panel or from capa — with its fujilib-calibration/1 record in the event's metadata.
unit_mismatch A channel is quarantined (above).
label_disagreement At the start of a run: the type code suggests another gas than channel_map asserts. A warning only.
setting_changed A setting written through capa during a run: the setting, its previous value and the new one.
write_uncertain / calibration_uncertain A write or calibration whose outcome could not be established.
metadata_stale The analyzer's settings could not be read; the snapshots show them as last read.

Events exist only during a run. A command from the manual control panel between runs is logged by the panel itself.

Capability flags

Flag Notes
READS_PROCESS_VAR The concentrations.
HAS_PARAMETER_CONFIG Settings writes: response times, ranges, hold, calibration gases.
HAS_GAS_CALIBRATION A zero or span driven from capa against a gas the operator supplies.
HAS_INTERNAL_CAL Added only when options asserts auto_calibration or auto_zero.
SUPPORTS_AUTO_RECONNECT Added when auto_reconnect = true.

Commands

All writes go through the authorization gate — issued_by plus either authorization_id or confirmed_by. On top of it the adapter applies a rule by fujilib's safety tier: everything DANGEROUS needs confirmed_by, a person at the interface, even inside an authorized run. A method step cannot change what the analyzer calibrates against.

Typed call DeviceCommand.kind Tier Needs
set_response_time(target, seconds) "set_response_time" persistent either authorization
set_output_hold(enabled) "set_output_hold" persistent either
set_hold_mode(mode) "set_hold_mode" persistent either
set_hold_value(channel, percent_fs) "set_hold_value" persistent either
set_range(channel, range_number) "set_range" persistent either
set_range_method(channel, method) "set_range_method" persistent either
— "write_parameter" the register's own by tier
— "apply_settings" the highest in the document by tier
set_calibration_gas(...) "set_calibration_gas" dangerous confirmed_by
return_to_measurement() "return_to_measurement" stateful either
— "calibration_plan" reads only either
begin_calibration(...) "calibration_begin" stateful confirmed_by
commit_calibration() "calibration_commit" dangerous confirmed_by
cancel_calibration() "calibration_cancel" stateful either
— "start_auto_calibration", "start_auto_zero_calibration" dangerous confirmed_by, and the option in options

Blowback is not exposed. Key lock, alarms and the calibration schedule are not writable.

A setting is written once and read back. apply_settings takes a fujilib-settings/1 document, compares all of it with the analyzer first, and writes nothing if any part is refused or is above the tier the command's authorization allows.

What comes back:

  • Accepted — the write read back as written (or the command did what it says). The detail names the change by gas and range span, as the setting_changed event does: O2 response time: 15 s -> 10 s, O2 range: 0–21 vol% -> 0–25 vol%, CO2 0–10 vol% span gas: 0.2 vol% -> 0.25 vol%. The event's metadata keeps fujilib's register name (response_time.o2).
  • Refused (accepted=False) — capa's gate, the tier rule, a bad payload, or fujilib or the analyzer declining: a value that does not fit, the analyzer calibrating or in a menu, an option not fitted, an exception reply. Nothing changed.
  • Uncertain (accepted=False, and an error event during a run) — the write did not read back as written, or its outcome is unknown. The analyzer may hold something other than what was asked for: read the setting back before going on.
  • Failed — the connection. The command raises.

Calibrating from capa

A zero or span is driven from the analyzer's card in the Manual Control dock. capa presses the analyzer's own calibration keys over Modbus; the operator switches the gas.

  1. Put the gas at the inlet. capa does not switch valves.
  2. Plan (optional) reads what the calibration would reach: every gas and range, and the calibration gas of each, e.g. span of O2: O2 0–21 vol% against 20.95 vol%. A zero can cover more than one gas when the analyzer is set to zero them together.
  3. Begin… — pick the gas, zero or span, and name the gas at the inlet: its value and a label (cylinder, lot) for the record. The unit is that of the range the gas measures on, and the analyzer's calibration-gas setting is shown beside the value: the gas named must equal it; change the setting first if it does not. After a confirmation, the panel is taken to its wait step. Nothing is calibrated yet.
  4. Wait for steady. The card shows whether the reading is steady on the named gas: it must move no more than 0.5 % of full scale over at least 30 s (longer for a long response time) and lie within 10 % of full scale of the gas.
  5. Hold to calibrate is enabled only while the reading is steady. Holding it sends the key that calibrates. fujilib checks everything again in the same breath — the panel, the gas, key lock, output hold, instrument errors — and refuses if anything has moved. A refused calibration leaves the run on the wait step.
  6. Cancel leaves the wait step without calibrating.

There is no way to calibrate on an unsteady reading from capa. If the reading will not settle within the rule, calibrate at the front panel: capa records that too (below).

What ends a run that nobody ended:

  • 15 minutes on the wait step — it cancels itself.
  • A run starting — it is cancelled, and that is the run's first event. The manual panel cannot reach a calibration during a run.
  • The hardware closing (config reload, exit) — the panel is returned to measurement first.
  • ESC at the front panel — the run ends without calibrating.

Every run that sent a key leaves a record: a fujilib-calibration/1 JSON document saved under configs/calibrations/analyzer/, named by serial number, channel, kind and UTC time, and never written over. It holds the plan, the named gas, the readings on the wait step, every key sent and how the panel was left.

A zero or span made at the front panel during a run is noticed by the adapter from the polls themselves: the rows are calibrating, and when it ends a calibration event carries its record. It costs no extra traffic.

Automatic calibration on a unit without valves

start_auto_calibration and start_auto_zero_calibration make the analyzer switch its own calibration-gas valves. On an analyzer whose gases are not plumbed through those contacts it would calibrate on whatever is at the inlet. The adapter refuses both unless options asserts the option, and the card does not offer them.

Discovery and handshake

discover(ports=None, addresses=(1,), timeout_s=0.3) wraps fujilib.find_devices: a Modbus RTU read to each station on each port, and an identification of whatever answers. Reads only, but every port scanned receives the probe frames, so the Setup tab scans serial families one at a time. One row per analyzer, keyed by port, address, model, serial, type_code, and the suggested channel_map.

handshake(params) is the per-device form capa validate --strict runs: read-only open + identify + close.

See Discovery for the cross-cutting UX.

Quirks

The channel map is asserted

Which gas a channel carries is not something the analyzer reports. Its type code suggests it, and the suggestion can be wrong or out of date: one analyzer capa was developed against carries O2 on a channel its type code says is not fitted. So channel_map is yours to assert, and it wins. A disagreement with the type code is a label_disagreement warning at the start of each run, nothing more.

Layer 2 validation flags a fuji_channel binding to a channel the map does not name (channels.fuji_channel_unmapped) and two analyzers on one port (devices.fuji.shared_port).

Warm-up and settling

When a connection failure is followed by a reconnect, every reading in the next 90 s that would be ok is settling instead — and valid is 0. A pulled USB cable may have taken the analyzer's power with it, and 90 s of good data marked is the cheaper mistake.

Two cases are not marked:

  • The first open. capa cannot know how long the analyzer has been on. Respect the manual's warm-up time.
  • A power cut that leaves the USB adapter powered. The port never fails, so nothing reconnects: the polls during the cut are error rows with a timeout, and the readings after it are ok while the analyzer warms up. A comm_lost / comm_restored pair whose error is a timeout rather than a connection failure is the sign; treat the minute or two after it as suspect.

One analyzer per port

fujilib addresses one station per open port. A second Fuji device on the same port fails validation rather than at open.

Unplugged between runs

The port is reopened by the stream, so an outage during a run heals itself. An adapter unplugged while no run is active stays disconnected until the next run starts or the config is reloaded; manual commands in between fail with a connection error.

A run that starts during an outage

device_records/fuji.parquet takes each column's type from its first 1,024 rows (or all of a shorter run). If every one of them is an error row, the reading columns are stored as text for that run. Nothing is lost, but a reader has to convert them.

Sim equivalent

capa.devices.sim.fuji_sim builds real fujilib frames, so its rows have exactly the real adapter's columns. It takes a channel_map (default CO2 / CO / O2 on CH1–CH3), a signal per channel (ambient air without one), tick_period_s, hold_from_s to switch output hold on part-way, and settle_s, how long a simulated calibration takes to read steady. The settings verbs change its own settings and a zero or span runs through the same three commands, so the card and declared device settings can be exercised offline. Its ranges are all in vol%: one per gas, two for O2 (0–25 and 0–10 vol%); a range change does not change the readings. It models no outages and no settling. See Simulators.