Destructive operations

Audience: operators issuing manual commands; plugin authors exposing controls in the Manual Control dock. Scope: commands flagged destructive=True and the two confirmation gates (modal dialog and hold-to-confirm button) that stand between an operator gesture and the device write.

A "destructive" command in CAPA terminology is one whose effect survives a power cycle, alters device state in a way the run does not need, or is dangerous to invoke accidentally. EEPROM writes, factory-style resets, valve closures, totalizer resets, and the Emergency Stop on the Run tab all qualify. The bar is survives power cycle OR can't be undone by issuing the opposite command.

The audit-trail guarantee from safety principle 1 applies just as it does to non-destructive writes — every confirmed destructive command lands in the bundle event log with issued_by and confirmed_by. The additional thing destructive commands get is a confirmation gate in the UI.


What counts as destructive today

The full list of destructive=True dispatches in the Manual Control cards, as of this writing:

Adapter Command kind What it does
Watlow set_setpoint (from the "Cool to safe" button) Drives the heater to a low setpoint; commits via the same write path as ordinary setpoints but is gated because a stuck low setpoint mid-experiment is destructive to the run. watlow.py:352
Watlow write_parameter Persistent parameter write to the controller — survives power-cycle. watlow.py:422
Sartorius balance internal_adjust Triggers the balance's motorized internal calibration cycle. balance.py:172
Sartorius balance save_menu Save menu settings to EEPROM. balance.py:234
Sartorius balance reload_menu Reload menu settings from EEPROM. balance.py:250
Alicat MFC set_gas (with save=True) Persist gas selection to EEPROM (the session-only variant is non-destructive). alicat.py:189
Alicat MFC hold_valves_closed Force the controller's valves fully closed regardless of setpoint. alicat.py:251
Alicat MFC totalizer_reset Zero a totalizer's accumulated flow history. alicat.py:280
Alicat MFC totalizer_reset_peak Zero a totalizer's peak-flow watermark. alicat.py:294
Fuji gas analyzer set_calibration_gas Changes the zero or span gas the analyzer calibrates a range against; the setting survives power-cycle and the next calibration is computed from it. fuji.py
Fuji gas analyzer calibration_begin Presses the analyzer's calibration keys up to its wait step. Nothing is calibrated yet, but the readings are marked calibrating until the operator calibrates or cancels. fuji.py
FLIR camera set_temperature_range Persistent camera-firmware config that survives power-cycle. camera.py:230

The key that actually calibrates the gas analyzer (calibration_commit) is not a destructive=True dispatch: it sits behind a hold-to-confirm button that is enabled only while the reading is steady on the named gas, and the adapter refuses it without a person's confirmation even inside an authorized run. See Fuji § Calibrating from capa.

Not on this list today: factory reset, baud rate change, and explicit gas-supply-line venting. If a future adapter adds those, flag them destructive=True and add a destructive_summary describing the consequence.


Gate 1: modal confirmation in manual cards

Every dispatch from a Manual Control card flows through ManualCardBase.dispatch(). The relevant snippet:

if destructive:
    summary = destructive_summary or f"{kind} on {self._name}"
    note = destructive_note or (
        "This may persist to EEPROM or otherwise alter device "
        "state in a way that survives power-cycle."
    )
    answer = QMessageBox.question(
        self,
        "Confirm device write",
        f"Confirm destructive operation:\n\n  {summary}\n\n"
        f"Operator: {operator}\n\n"
        f"{note}",
        QMessageBox.StandardButton.Yes | QMessageBox.StandardButton.No,
        QMessageBox.StandardButton.No,  # default is No
    )
    if answer != QMessageBox.StandardButton.Yes:
        self._set_status("cancelled", level="idle")
        return None

Four things to note:

  • The closing sentence can be replaced. A card passes destructive_note when the consequence is not a write that persists in the device — beginning a gas-analyzer calibration, for one — so the dialog says what the operation actually does.
  • Default is "No." The dialog opens with focus on No, so an accidental Enter-press cancels.
  • The operator id is shown in the dialog. This is the operator who will be stamped into the confirmed_by field if they click Yes.
  • Cancellation is logged as an idle status, not as an event. The card resets to its quiescent state; no confirmed row is written into the event log. (Future change: log cancellations to make accidental-near-misses auditable — track in your project notes.)

If the operator clicks Yes, the card constructs a one-off Authorization(operator_id, run_id="manual") and calls issue_manual() — see authorization gates.

The dialog is only shown when destructive=True. Non-destructive manual commands (zero_tare, set_setpoint to a normal value, totalizer read) dispatch immediately with no confirm dialog.


Gate 2: hold-to-confirm (Run tab Emergency Stop)

The Manual Control dialog handles "EEPROM survival" destructiveness. A separate widget — HoldToConfirmButton — handles "fast-clickable + lose-an-hour-of-work" destructiveness.

It is used in exactly one place today: the Run tab's Emergency Stop button. The mechanic:

  • Press and hold for HOLD_DURATION_MS = 1000 milliseconds.
  • A progress fill animates inside the button at ~30 FPS.
  • Release before the 1 s mark — the in-progress hold is silently cancelled and the button re-arms for the next press.
  • Hold for the full second — the confirmed signal fires, which (for the Emergency Stop) calls _on_abort_clicked("immediate").

Emergency stop is "immediate," not "graceful." The two buttons both stop the run and both produce a sealed bundle. They differ in the exit_reason stamped into the audit log — operator_immediate (Emergency Stop) vs operator_safe_shutdown (Stop run) — and in what the procedure does in response. The convention: operator_safe_shutdown means "honour your cleanup discipline (cool to safe target, run a SafeShutdownStep via run_segment, etc.) before unwinding"; operator_immediate means "exit as fast as you can." See shutdown sequence for the full mechanics — including the subtlety that the conductor itself does not differentiate, so the actual cleanup behaviour comes from the procedure.

The 1-second duration is a deliberate compromise: long enough that an accidental brush against the button cannot trigger an emergency stop on a 30-minute run, short enough that a deliberate operator response (rig misbehaving, operator wants to stop now) is not frustrating.

The widget is not themed to the manual-card visual language — it has its own destructive-red accent and font weight. That's intentional. Visually, the Emergency Stop should not be mistakable for a normal button.


Flagging a custom command destructive

If you're writing a custom manual card (e.g. as a plugin) and exposing a command that meets the bar — survives power cycle, can't be undone by issuing the opposite command, or has consequences disproportionate to a click — flag it.

The contract is: pass destructive=True and destructive_summary="…" to schedule_dispatch() (or the underlying dispatch() if you're driving directly). The base class handles the rest:

self.schedule_dispatch(
    kind="my_destructive_verb",
    payload={"foo": 1},
    destructive=True,
    destructive_summary=(
        "One sentence: what does this do, and what survives power-cycle? "
        "Mention EEPROM if relevant — operators look for that word."
    ),
)

The destructive_summary is what the operator sees in the confirmation dialog. Write it for an operator who knows the device but doesn't remember exactly which verb you're invoking. Mention EEPROM, flash wear, or "can't be undone" explicitly when applicable.

Do not plumb the hold-to-confirm widget into a Manual Control card. The widget is intentionally scoped to the Emergency Stop today — adding a layer of "hold for 1 s" on top of every destructive dispatch is more friction than the safety benefit warrants, and the modal dialog is the established convention.


What does NOT need the destructive gate

For symmetry, here are the kinds of writes that intentionally do not require destructive confirmation:

  • Procedure-issued setpoints during a run. They go through Authorization.issue() and inherit the run-arm coverage. The fact that the operator armed the run is the consent.
  • Method-step setpoints. Same — the method is part of the armed config.
  • Read commands. Any command whose effect is "ask the device for a value" — read_status, read_gas_list, get_totalizer, etc. — has no destructive component.
  • Session-only writes. E.g. Alicat set_gas without save=True is dispatched without confirmation — the gas selection reverts on power-cycle. Only the persistent save=True variant is destructive.

The principle: confirmation is for actions whose consequences the operator might not have anticipated. A setpoint change during an armed run is part of the contract the operator already agreed to.


See also: Authorization gates · Shutdown sequence · Manual controls · The Run tab