SHT4x ambient sensor

Audience: operators logging the lab's air temperature and relative humidity alongside a CAPA run. Scope: capa's capa.devices.sht4x adapter — [devices.params] fields, the wide-row emission, the sht4x_reading binding, the precision and heater command surface, and the bridge quirk the driver hides.


At a glance

Adapter id capa.devices.sht4x
Transport hidapi via the in-tree MCP2221 driver, capa.devices._mcp2221
Real adapter capa.devices.sht4x
Sim adapter capa.devices.sim.sht4x_sim
Resource scheme hid:mcp2221:<bridge serial or any>
Channel binding sht4x_reading
Emission shape wide_row — temperature and humidity together, one row per poll
Default poll rate 1 Hz; capped at 10 Hz

Supported hardware

A Sensirion SHT40, SHT41 or SHT45 (they share one protocol; the chip cannot say which it is) wired to the I2C pins of a Microchip MCP2221 or MCP2221A USB bridge. The bridge enumerates as a serial port plus a vendor HID interface; capa talks I2C over the HID interface and ignores the port. Unlike the serial families there is no sibling library: the whole sensor is six commands and a checksum, so the protocol lives in the adapter module and the bridge's four HID commands in a private module next to it.

Configuration

configs/hardware/sht4x_real.toml (excerpt)
[[devices]]
name = "ambient"
adapter = "capa.devices.sht4x"
on_failure = "warn"

[devices.params]
i2c_address = 0x44
precision = "high"      # "high" | "medium" | "low"
rate_hz = 1.0
i2c_speed_hz = 100000
auto_reconnect = true
snapshot_period_s = 30.0

Sht4xAdapterParams is extra="forbid".

Key Default Notes
serial unset USB serial of the MCP2221 to use. The chip ships with serial enumeration off, so with one bridge leave this unset and the first bridge found is used.
i2c_address 0x44 0x44 for SHT40-A, SHT41-A and SHT45; 0x45 for the -B variants; 0x46 for SHT40-C.
precision "high" Repeatability setting the stream starts with: an 8.3, 4.5 or 1.7 ms conversion. set_precision changes it live; device_settings can declare it.
rate_hz 1.0 Capped at 10 Hz: the datasheet keeps the measurement duty cycle under ten percent to limit self-heating.
i2c_speed_hz 100000 47 kHz to 400 kHz, the bridge's range.
auto_reconnect true A failed poll keeps its row, counts as a recoverable error and does not end the stream; a bridge I/O failure reopens the bridge after a one-second back-off. Toggles SUPPORTS_AUTO_RECONNECT.
snapshot_period_s 30.0 Health-ping cadence.

Set on_failure = "warn" on the device: ambient logging should degrade a run, not abort it.

Channels and emission shape

Every poll measures temperature and humidity together, so the adapter is a wide-row family: one row per poll into device_records/sht4x.parquet with temperature_c, humidity_pct, the raw 16-bit words temperature_raw and humidity_raw, the precision used, latency_ms from command to data, and error_type / error_message (both null on a good poll). The record's timestamp is the midpoint of the poll.

The only supported binding source is sht4x_reading, with field set to "temperature" or "humidity":

[[channels]]
name = "ambient.temperature"
kind = "temperature"
unit = "degC"
plot_group = "ambient"

[channels.source]
source = "sht4x_reading"
device = "ambient"
field = "temperature"

[channels.calibration]
kind = "identity"
input_unit = "degC"
output_unit = "degC"

[channels.metadata]
capa_group = "ambient_temperature"

The humidity channel is the same with field = "humidity", kind = "humidity", unit = "percent" and capa_group = "ambient_humidity". Both groups are optional under the CAPA profile. Humidity is clipped to 0–100 %RH as the datasheet asks; the raw word keeps any overshoot.

A failed poll (checksum mismatch, a sensor that did not answer, a bridge I/O error) preserves the native row with the error fields populated but emits no ChannelSample — the sample's status does not get to lie about a value that didn't arrive.

See channel bindings for the selector contract.

Capability flags

Flag Notes
HAS_PARAMETER_CONFIG Precision setting; the manual card.
HAS_SENSOR_HEATER The heater pulse command.
SUPPORTS_DISCOVERY capa hardware discover probes every bridge.
SUPPORTS_AUTO_RECONNECT Added when auto_reconnect=True.

Commands

Typed call DeviceCommand.kind Notes
set_precision(...) "set_precision" Payload {"precision": "high" \| "medium" \| "low"}. Changes which measure command the stream sends; nothing is written to the sensor.
heater_pulse(...) "heater_pulse" Payload {"power_mw": 200 \| 110 \| 20, "duration_s": 1.0 \| 0.1}. Runs the sensor's heater once and returns the high-precision reading taken at the end of the pulse.
soft_reset() "soft_reset" The sensor's reset command; the serial is re-read.
read_once() n/a Read-only — not gated. One measurement outside the stream.
read_state_snapshot() n/a Read-only — not gated. Precision, sensor serial, bridge revision, last reading and heater history, for the manual card.

All writes go through the authorization gate — issued_by plus either authorization_id or confirmed_by.

The heater

The SHT4x has a small heater meant for driving condensation off the element after a humid excursion, or for a plausibility check of the humidity response. It is not needed in normal CAPA use, and a pulse skews the reading warm and dry for a while after (several °C right after a short pulse, decaying over tens of seconds), so the card asks for confirmation and the stream's next records are not ambient. Two guards are enforced in the adapter, both answered with accepted=False rather than an error:

  • Ambient limit. A pulse is refused while the last reading is above 65 °C, the datasheet's limit for heater use.
  • Duty cycle. Heater-on time is held to ten percent: after a 1 s pulse the next is allowed 9 s later, after a 0.1 s pulse 0.9 s later.

Each pulse is logged as a heater_pulse event with the end-of-pulse reading in its metadata, so a skewed stretch of the record is explained in events.sqlite.

An experiment can declare the precision under device_settings: to have it applied when the config loads; the heater is never declarative.

Discovery and handshake

discover() enumerates every MCP2221 on the machine and probes each for a sensor at 0x44, 0x45 and 0x46, returning one row per sensor found: the bridge's serial (null unless enumeration is enabled in its flash) and HID path, the i2c_address and the sensor_serial. Bridges another process holds are skipped.

handshake(params) is the per-device form capa validate --strict runs: open, soft reset, read the serial and one measurement, close.

See Discovery for the cross-cutting UX.

Quirks

The bridge sticks after a NACK

The sensor NACKs its address while a conversion is in progress, and the MCP2221 answers that by parking its I2C engine in an error state in which every later transfer is refused until a cancel command. The driver cancels on every NACK and on open, so a read issued a little early, a process killed mid-transfer or a sensor unplugged and replugged all recover without operator action. The adapter also waits the datasheet's maximum conversion time before reading and keeps retrying a NACKed read for 20 ms after that, so a healthy sensor never surfaces this at all.

One sensor per bridge

The adapter opens the bridge itself, so a second SHT4x on the same bus (the 0x44 and 0x45 variants can share one) is not supported yet; declare one device per bridge. Two bridges are told apart by serial once serial enumeration is enabled in their flash.

Placement

The sensor measures the air around it, including whatever the cone heater and the PC radiate. Mount it away from the heater's plume and the exhaust, and expect it to read a degree or so above the room if it sits on a warm enclosure.

Sim equivalent

capa.devices.sim.sht4x_sim takes temperature_signal and humidity_signal specs (both optional; the defaults are a steady 22 °C and 50 %RH), tick_period_s and precision. It keeps the precision and heater state the real adapter keeps, answers read_state_snapshot() the same way, and a heater pulse adds a warm offset that decays over ten seconds so the perturbation shows up in a sim run.